Question
I have set my VNA to single sweep mode and switched on averaging with a factor of 10.
But when I restart the single sweep, it stops after 1 sweep and the averaging is not done.
I have set my VNA to single sweep mode and switched on averaging with a factor of 10.
But when I restart the single sweep, it stops after 1 sweep and the averaging is not done.
You also have to set Define Restart in the right side menu of the Sweep menu.
In this menu point, change from
Single Sweep
to
Single Group of 10 Sweeps
10 sweeps because you have an averaging factor of 10.