Question
I have a UPV and I can see in the Windows control panel under System and Device Manager that the settings for
COM1 are 9600,8,N,1.
I use these settings but no command is accepted.
I have a UPV and I can see in the Windows control panel under System and Device Manager that the settings for
COM1 are 9600,8,N,1.
I use these settings but no command is accepted.
The settings for remote control of the UPV via RS-232-C must be as follows:
Baud rate 19200
Data bits 8
Parity None
Stop bits 1
Flow control Hardware