Vertical system
|
Analog channels |
|
4 |
Digital channels (opt.) |
|
16 |
Analog bandwidth (–3 dB) |
R&S®MXO 4 with (R&S®MXO 4 -B24x options) |
200 MHz, 350 MHz, 500 MHz, 1 GHz, 1.5 GHz |
Rise time |
R&S®MXO 4 (with R&S®MXO 4 -B24x options) |
1.75 ns, 1 ns, 700 ps, 350 ps, 234 ps |
Input impedance |
|
50 Ω ± 1.5 %,
1 MΩ ± 1 % || 12 pF (meas.) |
Input sensitivity |
at 50 Ω |
0.5 mV/div to 1 V/div,
entire analog bandwidth supported for all input
sensitivities |
|
at 1 MΩ |
0.5 mV/div to 10 V/div,
entire analog bandwidth supported for all input sensitivities |
Vertical resolution |
|
up to 18 bit with HD mode (high definition) |
DC gain accuracy |
offset and position set to 0 V, after self-alignment |
|
|
input sensitivity > 5 mV/div |
±1 % full scale |
|
input sensitivity ≤ 5 mV/div to ≥ 1 mV/div |
±1.5 % full scale |
|
input sensitivity < 1 mV/div |
±2.5 % full scale |
Acquisition system
|
Sampling rate |
analog channels (real time) |
max. 5 Gsample/s on 2 channels,
max. 2.5 Gsample/s on 4 channels |
|
analog channels (interpolated) |
max. 5 Tsample/s |
|
digital channels |
max. 5 Gsample/s on each channel |
Acquisition memory |
standard |
400 Mpoints with 4 active channel (single),
400 Mpoints with 2 active channel (run) |
|
R&S®MXO4-B108 option |
800 Mpoints with 2 active channel (single),
800 Mpoints with 1 active channel (run) |
Max. acquisition rate |
|
> 4 500 000 waveforms/s |
Acquisition modes |
|
sample, peak detect, average, envelope |
Interpolation modes |
|
linear, sin(x)/x, sample & hold |
Horizontal system
|
Timebase range |
|
Selectable between 200 ps/div and 10 000 s/div
|
Timebase accuracy |
after delivery/calibration, at +23 °C |
±0.2 ppm |
|
During calibration interval |
±1 ppm |
Channel-to-channel skew |
between analog channels |
< 100 ps (meas.) |
|
between digital channels |
< 500 ps (meas.) |
Trigger system
|
Standard trigger types |
|
Edge, glitch, width, runt, window, timeout, interval, slew rate, setup and hold, pattern, state, zone trigger |
Advanced trigger modes |
|
A/B/R, serial bus, trigger in and out |
Trigger sensitivity |
|
0.0001 div, from DC to instrument bandwidth for all
vertical scales |
Analysis and measurement
|
Automatic measurements |
|
39 |
Cursor measurements |
|
4 |
Waveform mathematics |
|
Up to 5 equations math on math (add, subtract, multiply, divide, absolute value,
square, square root, integrate, differentiate,
log10, loge, log2, reciprocal, invert, low pass,
high pass, rescale (a*x+b)) |
Logic analyzer (MSO) functionality
|
Digital input channels |
|
16 |
MSO Bandwidth |
|
400 MHz |
MSO sampling rate |
|
5 Gsamples/s per channel |
General data
|
Dimensions |
W × H × D |
414 mm × 279 mm × 162 mm
(16.3 in × 10.99 in × 6.38 in) |
Weight |
without options, nominal |
6.0 kg (13.23 lb) |
Interfaces |
|
1 x USB 3.0 super speed port
2 x USB 2.0 high speed port
2 x USB 3.1 gen1 ports
1 x USB 3.1 gen1, type B plug
1 Gbps LAN
HDMI for external monitor
External trigger, trigger output
Ref In, ref out
VESA mount |
Display |
type |
13.3" LC TFT color display with capacitive
touchscreen |